RADIATION DOSE AND IMAGE QUALITY FOR ADULT INTERVENTIONAL CARDIOLOGY IN CHILE: A NATIONAL SURVEY
No Thumbnail Available
Date
2011
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
The aim of this work was to investigate the differences in dose settings and image quality among 10 X-ray systems used for interventional cardiology in Chile. Entrance surface air kerma (ESAK) was measured on a phantom of 20 cm thickness of polymethyl methacrylate slabs. Image quality was evaluated using DICOM images of a test object Leeds TOR 18-FG for cine mode acquisition, through the numerical parameters signal-to-noise ratio (SNR), high-contrast spatial resolution (HCSR) and figure of merit. ESAK rate values for fluoroscopy modes ranged between 7.1 and 121.7 mGy min(-1). For cine mode, ESAK values per frame ranged from 63 to 400 mu Gy fr(-1). SNR and HCSR parameters for cine mode varied from 4.8 to 8.6 and 0.4 to 10, respectively. FOM values resulted from 6.9 to 64.5 among the different X-ray systems. Results show important differences between systems and point out the need to launch an optimisation programme.